Option Menu #:
32
5-32 OPTION MENU #:32 DIAL-UP TESTLINE
OPTION MENU #:32, DIAL-UP TESTLINE, is a standard feature that enables the 930i to provide Quiet
Termination or send a tone in response to ringing or seizure. That is, the 930i can act as a Type 100
or a Type 102 testline. Note that, since the tone will not turn off, it is not a true 102 testline (i.e., a noise
reading without tone cannot be made).
To set up the 930i as a Type 100 or 102 testline:
1. Use the Send Tone function to setup the tone to be sent. If the Send Tone function is turned
off, the 930i will supply Quiet Termination.
2. Press the Option Menu function key.
3. Press the Up/Down Arrow keys to scroll through the option menus, OR select 32 using the
numeric keypad and then press the Option Menu function key. The display reads:
OPTION MENU #: 32 DIAL-UP TESTLINE
K1
K2
K3
K4
4. Press ENT or any soft key (K1-K4) to activate your selection. The display briefly reads:
WAITING FOR SEIZURE EXIT
K1
K2
K3
K4
You will then advance to the initial Dial/Ring display to set up and make your call. This display
will differ depending on whether you are in analog or PCM mode. Refer to
Section 3-2.2.1
for
a description of the Dial/Ring function in analog, and
Section 3-2.2.2
for a description of Dial/
Ring in PCM.
Note:
Before receiving a call in PCM, you must select and set up a protocol if one has not already been
selected under the Dial/Ring function key. Refer to
Section 3-2.2
for a complete description of
this procedure. After setup is completed, you will be ready to receive a call.
Entering this Option Menu places the 930i on-hook where it will remain waiting for line seizure before
going off-hook.
If the 930i tone generator has been turned OFF, the 930i will provide Quiet Termination when ringing
or seizure is detected.
If you have selected a tone, for example, 1020 Hz at 0 dBm, and left the tone generator turned on prior
to entering OPTION MENU #:32, then the 930i will respond with the 1020 Hz at 0 dBm when ringing
or seizure is detected. It will remain until seizure is dropped (on-hook), the Option Menu function key
is pressed, or a 3 minute time out has occurred.
This option has a 3 minute time out if forward disconnect is not available.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 930i
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 203: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 235: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...