6 Acquiring Images
70
Basic User Manual
6. Optimize the PW-mode image. For details, refer to Section 6.5.3 Optimizing Spectral Doppler Images.
If needed, select the
Update
key on the control panel to enter the inactivated B+PW mode 2, the following
screen appears.
0
5
50
-50
cm/s
X-axis: Time
Y-axis:
Frequency
s
Figure 6-13
Inactivated B+PW Mode 2 Imaging Screen
You can also adjust the position and angle of spectral Doppler line, the size of sample volume gate, and the
direction of the flow cursor
7. Select the
PW
key to exit.
6�5�2
CW Mode
Continuous Wave Doppler (CW) uses continuously transmitted and received ultrasound energy to generate a
spectral display. CW Doppler is used to measure a high velocity, the highest velocity at a specific site, but no
aliasing is created.
Perform the following steps to acquire CW-mode images.
1. Optimize a B-mode image.
2. Select the
CW
key on the control panel to enter the inactivated B+CW mode 1.
0
5
Spectral Doppler Line
Flow Cursor
Sample Volume Gate
2D Imaging
S
Figure 6-14
Inactivated B+CW-Mode 1 Imaging Screen
−
The spectral Doppler line and the sample volume gate are used to locate the qualitative analysis on the
image.
−
The flow cursor needs to be adjusted parallel to the flow when measuring the velocity.
3. Adjust the position and angle of the spectral Doppler line.
−
Position the sample volume gate on the spectral Doppler line by moving the trackball up or down.
Summary of Contents for EVUS 8
Page 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Page 10: ...This page is intentionally left blank...
Page 18: ...This page is intentionally left blank...
Page 62: ...This page is intentionally left blank...
Page 88: ...This page is intentionally left blank...
Page 92: ...This page is intentionally left blank...
Page 112: ...This page is intentionally left blank...
Page 122: ...This page is intentionally left blank...
Page 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...