27
4
Connection Diagnostics
This procedure presumes the operator has basic knowledge of the ROWETECH HASCP and has
experience with serial communication interfaces and test equipment. For further assistance, contact Rowe
Technologies directly.
Connect HASCP communication port(s) to the Windows PC (outlined above). Open the ROWETECH
HASCP software. Select communication port and Baud rate.
Select the Terminal screen on the HASCP software and turn on the power supply to the HASCP and click
Break <CR>.
If you see the Wake Up message on the Terminal screen:
Copyright (c) 2019 Rowe Technologies Inc. All rights reserved.
HASCP
DP1200 DP600 DP2400 DP1200
SN: 08BJHQ00000000000000000000888888
FW: 00.07.27 Nov 20 2019 05:22:14
STOP+
If there is no Wakeup message displayed on the Terminal screen:
•
Verify the Baud rate and Communication port setting.
•
Verify the wiring of the communication port.
•
Verify RS232 communications on the PC by connecting pins 2 and 3 together on the DB9
connector going into the PC. Then send a command for e.g. START to the port. You should see
the command/characters echoed on the Terminal screen.
•
Try a NULL MODEM on the RS232.
•
Connect an Oscilloscope to the RS485 data lines. Then send a command for e.g. START. You
should see the command/characters on the scope display the bit rate should match the PC Baud
rate.
•
Try swapping the 2 data lines on the RS485.
•
Connect an oscilloscope to the ADCP/DVL output data lines on either communication port.
•
Turn OFF the HASCP power. Wait 10 seconds. Turn ON the HASCP power supply:
Measure the voltage at the input to the HASCP pin1 on the power connector. The voltage should be
between 12 and 24 VDC depending on the power supply. If the voltage is out of range find another
supply.
Measure the power supply current. The power up current should be 0.2 to 0.5 Amps if the system is
awake and waiting for a command. If the system is pinging it will be drawing 2 to 3 Amps and 0.2 to 0.3
Amps the rest of the time.
Remove the underwater housing keeping the wiring connectors attached. If there is still an issue contact
ROWETECH with the results of these tests.
Summary of Contents for HASCP
Page 26: ...26 Figure 15 Firmware Tab in the RTI HASCP software...
Page 68: ...68 8 Cable Wiring Diagrams This section outlines the HASCP communication cable wiring...
Page 73: ...73 10 Mechanical Drawings and Assembly Figure 23 Outline drawing of the HASCP unit MCBH 10M...
Page 74: ...74 Figure 24 Location of transducer beams in the HASCP unit...
Page 76: ...76 Figure 26 Exploded view of the HASCP unit showing the details of the hardware used...
Page 78: ...78 Figure 28 HASCP mounted on a pole used for internal testing at RTI...