Probe
Probe
Probe
Probe 1
1
1
1
(applies
to
HygroClip 2
probe
only)
( 仅应 用于
HygroClip 2
探头)
Info
信息
Serial Nbr
序列号
Version
版本
Firmware version
固件版本
Name
名称
User defined (HW4)
用户定义(
HW4
)
Humi Adjust
Acquired
Points
已
保存的校
准点
Lists the cal.
points present in
the probe
memory (a max.
of 10 points are
shown)
探头存储
器内当前的校准
点列表(最多可见
10
个点)
Additional Options:
附加功能:
o Clear all cal. points
清除所有校准点
o Clear the last point
清除最后一个校准点
Acquire
Ref
参考
值
Saves value measured by probe 2 as a cal.
Point
探头
2
保存测量值作为校准点
Acquire
获取
Manual entry:
known reference
environment
手
动进入:已知参考
环境
Saves value entered manually as a cal.
Point
保存手动键入的值作为一个校准点
Adjust
Adjusts the probe
调节探头
Effect depends on number of calibration
points in probe memory
取决于探头存储器
内保存的校准点的数量
Reset to
Factory
恢复出厂
设置
Returns the probe to the initial factory
adjustment
恢复探头最初出厂调整值
Temp Adjust
Acquired
Points
Lists the cal.
points present in
the probe
memory (a max.
of 2 points are
shown)
探头存储
器内当前的校准
点列表(最多可见
10
个点)
Additional Options:
附加选项:
o Clear all cal. Points
清除所有校
准点
o Clear the last point
清除最后的一个校准点
Acquire
Ref
Saves value measured by probe 2 as a cal.
point
探头
2
保存测量值作为校准点
Acquire
Manual entry:
known reference
environment
手动进入:已知参
Saves value entered manually as a cal.
point