R&S FSU
Firmware Release 4.61 SP3
7
E-33
New Functions in Version 4.61
Auto Login Password for user INSTRUMENT is changed to "894129" for security reasons
Support for Noise Correction outside of ACP measurement
Multi Carrier ACP with up to 18 TX Channels
Multi Carrier ACP: Support for save/recall of user defined standards
Spurious Emissions Measurement: Supports for save/recall of user defined standards
SEM measurement: Supports for save/recall of user defined standards
SEM measurement: Ref Level dialog available to adjust the sweep list's level settings
SEM measurement: Additional WIMAX configuration files available for Downlink ETSI
(5MHz / 10MHz)
Extended Marker Peak List function includes automatic peak list update
FS-K7: New Fundamental Frequency AUTO/MANUAL setting for SINAD and THD
measurement
FS-K9: Indication of the power meter's serial number
Resolution Bandwidth 6.25 kHz supported (with Service Pack 2)
FS-K9: Support for Power Sensor NRP-Z86 available (with Service Pack 2).
PSA / 89600 Emulation available (with Service Pack 2)
Support for the Status Operation Register Bits MEASuring/SWEeping (with Service Pack 2)
Improvements
The version numbers in brackets indicate the version in which the issue was observed for the first time.
1. (V4.41) The remote command "MMEM:DATA" ignores the default folder, set by "MMEM:CDIR".
The following command sequence should store data in the file 'filenname' in folder d:\user\settings:
MMEM:MDIR 'd:\\user\\settings'
MMEM:CDIR 'd:\\user\\settings';*opc?
MMEM:DATA 'filename',#<Blockdaten>
The file is created in folder "d:\r_s.fw\instr" instead. The command to change the path had no effect.
Using a filename with path included, works fine, e.g.
MMEM:DATA 'd:\\user\\settings\\filename',#<Blockdaten>.
Note: This issue is already fixed in V4.51 SP1
2. (V4.51) Option Key disabled after reboot.
The option key was manually enabled in the SETUP – GENERAL SETUP – OPTIONS list. After reboot of
the instrument, the option is disabled again.
Note: This issue is already fixed in V4.51 SP1
3. (V4.51) The operating hour is set to 100.000 hours.
Note: This issue is already fixed in V4.51 SP1
4. (V4.51) Spurious Signal at 758.12 MHz eliminated (with Center Frequency 620 MHz).
The spurious appears with a Center Frequency of 620 MHz, a Resolution Bandwidth of 50 kHz, a Video
Bandwidth of 300 kHz, a Sweeptime of 5 sec and 10001 Sweep Points.