R&S FSU
Firmware Release 4.61 SP3
18
E-33
Remote Control – Description of the Status Registers
STATus:OPERation Register
In the CONDition part, this register contains information on which actions the instrument is being executing
or, in the EVENt part, information on which actions the instrument has executed since the last reading. It
can be read using commands
"STATus:OPERation:CONDition?
" or
"STATus:OPERation[:EVENt]?"
.
Bit No.
Meaning
0
CALibrating
This bit is set as long as the instrument is performing a calibration.
1 to 2
These bits are not used
3
SWEeping
This bit is set while the instrument performs a sweep.
It is supported in analyzer mode only (Full Screen, frequency domain and time domain).
4
MEASuring
This bit is set while the instrument performs a measurement.
It is supported in analyzer mode only (Full Screen, frequency domain and time domain).
5 to 7
These bits are not used
8
HardCOPy in progress
This bit is set while the instrument is printing a hardcopy.
9
This bit is not used
10
Sweep Break
This bit is set when end of sweep range is reached (spurious measurement, mode analyzer).
Command “INIT:CONM” has to be used to proceed.
11 to 14
These bits are not used
15
This bit is always 0