:[SENSe<1|2>:]PMETer:DCYCle:VALue
0.001 … 99.999
This command sets the duty cycle to a percent value for the correction of pulsemodulated signals. With
the correction activated (SENS:PMET:DCYC:STAT ON), the sensor calculates the signal pulse power
from this value and the mean power. Valid entries are from 0.001% to 99.999%; the stepsize is 0.1%; the
maximum resolution for numerial entries is 0.001%. The default setting is 99.999%
Example:
":SENS:PMET:STAT ON"
' activate power meter
":SENS:PMET:DCYC:STAT ON"
' switch the correction on
":SENS:PMET:DCYC:VAL 50.0"
' set the duty cycle to 50 %
Properties:
*RST value:
99.999 PCT
SCPI:
device-specific
R&S FS-K15 Extensions
The R&S FS-K15 VOR/ILS Avionics Measurements Application functions are included in a separate manual
set. Please refer to the following order numbers:
Operating Manual "VOR/ILS Avionics Measurements Application Firmware R&S FS-K15":
•
1302.0942.42-02 (English)
The corresponding PDF-File ist separately available on the service board.
Last minute changes to the R&S FS-K15 operating manual
None.