![Rohde & Schwarz FSH13 Operating Manual Download Page 327](http://html.mh-extra.com/html/rohde-and-schwarz/fsh13/fsh13_operating-manual_1477832327.webp)
R&S FSH
Digital Modulation Analyzer
Measurements on LTE Signals
Operating Manual 1173.6275.12 - 30
312
Frame Offset
Shows the time difference between the (sub)frame start and capture buffer start.
The frame offset is only shown when you turn off the occupied bandwidth
measurement.
Occupied Bandwidth
Shows the occupied bandwidth of the signal in Hz.
You can control if the occupied bandwidth is measured or not. By default, it is not.
► Press the MEAS key.
► Press the "Meas Settings" softkey.
► Select the "[ ] Occupied Bandwidth" menu item.
The R&S FSH determines the occupied bandwidth in addition to the other global
results. When the measurement has been turned on, the menu item is marked by
an [X].
Cell Identity [Grp/ID]
Shows the cell identity, cell group and cell ID the current results correspond to.
For more information see "
Cyclic Prefix
Shows the cyclic prefix the current results correspond to.
For more information see "
Traffic Activity
Percentage of traffic slots with data.
SINR
Shows the Signal to Interference and Noise ratio.
The SINR is the ratio of the signal power and the sum of interference and noise power.
RSSI (3GPP TS 36.214)
Shows the Received Signal Strength Indicator (RSSI).
The RSSI is the complete signal power of the channel that has been measured,
regardless of the origin of the signal.
RSRQ (3GPP TS 36.214)
Shows the Reference Signal Received Quality (RSRQ).
The RSRQ is the ratio of the RSRP and the RSSI.