R&S FSH
Spectrum Analyzer Mode
Using Transducer Factors
Operating Manual 1173.6275.12 - 30
132
3.4.3 Frequency Range of Transducer
If the set frequency range is wider than the span in which a transducer is defined, the
R&S FSH assumes the transducer values outside the defined range to be zero.
3.4.4 Data Sets Containing Transducer Factors
The R&S FSH stores data sets together with any transducer factors that may have
been active for the measurement in question. When such a dataset is recalled, the
associated transducer factor(s) are switched on as well. Transducer factors recalled as
part of a data set do however not appear in the list of transducer factors.