9.6 Measurements on 1xEV-DO Signals
Equipped with the software application R&S FSH-K47, you can perform measurements
on downlink 1xEV-DO signals in accordance to the 3GPP2 standard with your
R&S FSH.
► Press the MODE key
► Press the "Dig Mod Analyzer" softkey.
► Select the "1xEVDO BTS" menu item.
The R&S FSH starts the signal analysis.
After you have started the measurement, the R&S FSH first records a section of the
signal that lasts approximately 27 ms (or about 1 Sync Frame Period). If the instrument
is in a triggered mode (external or GPS trigger), it then searches for the start of the PN
code in the vicinity of the PN offset you have entered. If you are using the automatic
PN Offset, the R&S FSH performs a full search for the PN offset and reports the PN
found. For a "free run" situation, the R&S FSH searches for the start of the PN
sequence over the entire signal length.
The R&S FSH performs various general measurements over one slot as well as
measurements on specific channels. It then shows the results in tabular form in the
"Result Summary" display,
To get highest measurement accuracy, it is necessary to synchronize the reference
frequency of the R&S FSH with the base station via the EXT REF IN input or with an
optional GPS receiver.