![Rohde & Schwarz FSH13 Operating Manual Download Page 278](http://html.mh-extra.com/html/rohde-and-schwarz/fsh13/fsh13_operating-manual_1477832278.webp)
R&S FSH
Digital Modulation Analyzer
Measurements on 3GPP FDD Signals
Operating Manual 1173.6275.12 - 30
263
I/Q Imbalance
Shows the I/Q Imbalance of the signal in %.
This value is valid only when the R&S FSH is in channel search mode.
Composite EVM
Shows the Error Vector Magnitude in %.
The EVM is defined as the ratio of the mean error power of the signal to the power of
an ideal reference signal. To calculate the mean error power, the root mean square
average (of the real and imaginary parts of the signal) is used.
This value is valid only when the R&S FSH is in channel search mode.
Active Channels
Shows the number of active channels in the signal.
Scr Code Found
Shows the number of the primary and secondary scrambling code, regardless if it has
been found automatically or entered manually.
Occupied Bandwidth
Shows the occupied bandwidth of the signal in Hz.
You can control if the occupied bandwidth is measured or not. By default, it is not.
► Press the MEAS key.
► Press the "Meas Settings" softkey.
► Select the "[ ] Occupied Bandwidth" menu item.
The R&S FSH determines the occupied bandwidth in addition to the other global
results. When the measurement has been turned on, the menu item is marked by
an [X].
Peak CDE (15 ksps)
Shows the Peak Code Domain Error of the signal in dB.
The Peak Code Domain Error is defined as the maximum code domain error power
that occurs in the measurement. The code domain error is the difference in power of
the test signal and an ideal reference signal.
This value is valid only when the R&S FSH is in channel search mode.
Avg. RCDE (64QAM)
Shows the Average Relative Code Domain Error of the signal.
Note that only channels with a 64QAM modulation are considered in this
measurement.