background image

3

EN

EN

2

SAFETY WARNINGS

The following safety information must be observed to insure maximum personal safety 

during the operation at this meter:

Measurements beyond the maximum selected range must not be attempted.

Extreme care must be taken when measuring above 50 V, especially on live

bus-bars.

To measure voltage, the instrument must not be switched to a current or resistance

range, or to the diode check or buzzer position.

Circuits 

must be de-energised and isolated before carrying out resistance tests.

The rotary selector switch must only be turned after removing test connections.

All external voltages must be disconnected from the instrument before removing

the battery.

Test leads and prods must be in good order, clean, and with no broken or

cracked insulation.

UK Safety Authorities recommend the use of fused test leads when measuring voltage

on high energy systems.

Replacement fuses must be of the correct type and rating.

The instrument must not be used if any part of it is damaged.

Warnings and precautions must be read and understood before an instrument is

used. They must be observed during the operation of this instrument.

Symbols used on this instrument are:

          Caution: refer to accompanying notes.

This symbol indicates that the operator must refer to an explanation in the  

Operating Instructions to avoid personal injury or damage to the meter.

Caution: risk of electric shock 

This WARNING symbol indicates a potentially hazardous situation, which if not  

avoided, could result in death or serious injury.

This CAUTION symbol indicates a potentially hazardous situation, which if not  

avoided, may result damage to the product.

This symbol advises the user that the terminal(s) so marked must not be    

connected to a circuit point at which the voltage with respect to earth ground  

exceeds (in this case) 1000 VAC or VDC.

Equipment protected throughout by Double Insulation (Class II)

Equipment complies with current EU directives.

SYMBOLS AND ANNUNCIATORS

Continuity

Low 

Battery

Diode test

Data 

Hold

AutoRanging

Alternating Current or Voltage

Direct Current or Voltage

Volts

Current range

OPERATION

To turn on the instrument turn the range knob from the OFF position to any  

measurement range.

Note:

For best battery life ALWAYS turn the function switch to the OFF position when the meter is 

not in use. This meter has Auto OFF that automatically shuts the meter OFF if 30 minutes 

elapse between uses.

NOTE: On some low AC and DC voltage ranges, with the test leads not connected to a 

device, the display may show a random, changing reading. This is normal and is caused by 

the high-input sensitivity. The reading will stabilize and give a proper measurement when 

connected to a circuit.

Summary of Contents for 355-00011

Page 1: ...www rnd electronics com Autoranging Digital Multimeter User Manual RND 355 00011...

Page 2: ...ct This symbol advises the user that the terminal s so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds in this case 1000 VAC or VDC Equipment...

Page 3: ...ition 3 Use the MODE button to select AC or DC Voltage 4 Connect the test leads in parallel to the circuit under test 5 Read the voltage measurement on the LCD display DC CURRENT MEASUREMENT 1 Insert...

Page 4: ...on can be evaluated as follows A If one reading shows a value and the other reading shows OL the diode is good B If both readings show OL the device is open C If both readings are very small or zero t...

Page 5: ...ication Low battery indication A battery symbol is displayed when the battery voltage drops below the operating level Measurement rate 2 times per second nominal Auto power off Meter automatically shu...

Page 6: ...range 10A dc on 10A range AC Current Auto ranging Overload Protection FF500mA 1000V and F10A 1000V Fuse AC Response 50 Hz to 400 Hz Maximum Input 400uA ac rms on uA 400mA ac rms on mA 10A ac rms on 1...

Page 7: ...below the operating voltage the battery warning symbol will appear in the LCD display The battery should be replaced 2 Follow instructions for installing battery See the Battery Installation section o...

Page 8: ...om its holder by gently pulling it out 4 Install the new fuse into the holder 5 Always use a fuse of the proper size and value 0 5A 1000V fast blow for the 400mA range 10A 1000V fast blow for the 10A...

Page 9: ...nleitung beziehen muss um Verletzungen oder Sch den am Messger t zu vermeiden Achtung Stromschlaggefahr Dieses WARNUNG Symbol weist auf eine potenziell gef hrliche Situation hin die zu schweren oder t...

Page 10: ...pannung aus 4 Schliessen Sie die Messleitungen parallel an den zu pr fenden Stromkreis an 5 Lesen Sie die Spannungsmessung auf der LCD Anzeige ab DC STROMMESSUNG 1 Stecken Sie den schwarzen Bananenste...

Page 11: ...er Messung ein Wert und bei der anderen OL angezeigt wird ist die Dio de in Ordnung B Wenn bei beiden Messungen OL angezeigt wird ist das Ger t defekt sperrt in beide Richtungen C Wenn die beiden Mess...

Page 12: ...Das Messger t schaltet sich nach ca 30 Minuten Inaktivit t automatisch ab Betriebsumgebung 10 C bis 50 C 14 F bis 122 F bei 70 relativer Luftfeuchte Lagertemperatur 30 C bis 60 C 4 F bis 140 F bei 80...

Page 13: ...0 des Messwerts 5 Stellen berlastschutz FF500 mA 1000 V und F10 A 1000 V Sicherung AC Reaktion 50 Hz bis 400 Hz Maximaler Eingang 400 A AC eff auf A 400 mA AC eff auf mA 10 A AC eff im 10 A Bereich Wi...

Page 14: ...ICHERUNG austauschen WARNUNG Um einen Stromschlag zu vermeiden trennen Sie die Messleitungen von einer Spannungsquelle bevor Sie die Batteriefachabdeckung entfernen 1 Wenn die Batterien leer sind oder...

Page 15: ...endreher l sen 3 Entfernen Sie die alte Sicherung aus ihrer Halterung indem Sie sie vorsichtig herausziehen 4 Setzen Sie die neue Sicherung in den Halter ein 5 Verwenden Sie immer eine Sicherung der r...

Page 16: ...3 DE DE...

Page 17: ...www rnd electronics com...

Reviews: