Changing and calling up the programme data
Parameters of the output functions
17
SYCON
3000 C
Output functions: Permanent signal 2
Activation with a limit overrun or underrun and in the event of a fault.
Automatic delete function 2
S t e p n o . : 6 . 5
A u t o m a t . d e l e t e 2 Y / N
▲
▼
►
▲
You can programme whether further analyses are carried out after the activa-
tion of the output relay due to the limit over-/underrun or the faults pro-
grammed in programme step 6.7. If further analyses are conducted, the relay is
automatically deactivated again if the result "Water good" occurs in one of the
next analyses.
If no further analyses are conducted, the relay must be deleted by hand or via
the input "Delete relay". Moreover, an analysis must be restarted.
Automatic deletion Yes: Consecutive analyses even after a limit
over/underrun or fault
Automatic deletion No:
Analysis stop in the event of a limit over/underrun
or fault
Number of the bad messages
S t e p n o . : 6 . 6
N u m b e r m e s s a g e s 2 *
▼
#
▲
You can programme the number of bad messages (limit overruns/underruns)
after which the output function is activated. Values from 1 to 5 can be entered.
With a value greater than 1, it is laid down in programme step 6.9 or 6.10
when the next analysis occurs.
Activation through a fault
S t e p n o . : 6 . 7
R F | N I | B V | V
H | V L |
F a u l t f i l l i n d i A C T I V
▼
►
#
As opposed to the output function "Permanent signal 1", you can determine with
the output function "Permanent signal 2" which faults cause the relay to be acti-
vated in addition to the limit over/underrun.
Using the "#" key, choose between the display "-" and "|".
"-" = No activation through displayed fault
"|" = Activation through displayed fault
RF
= Refill indicator
VH
= Measured value too high
NI
= No indicator (Stop)
VL
= Measured value too low
BV
= Blank value too low
Output function: Analysis is running
Analysis delay
S t e p n o . : 6 . 8
A n a l y s i s d e l a y 5 s
▲
▼
►
#
The start of an analysis – opening of the input valve – can be delayed within the
range of 0-999 seconds.
6.9
Output
Bad messages
6.6
*
Permanent
signal 2
Output
Analysis runs
6.2
no
yes
no
?
yes
6.9
6.4
?
Fault messages
6.7
RF
NI
BV
VH
VL
Autom.delete 2
6.5
Yes
No
Delay
6.8
s