Notes on connections
Connection and programming examples
28
SYCON
3000 C
Recorder interface
The analyser is equipped with 2 analogue outputs. The various instrument
states can be extracted as analogue values between 0-20 mA at the output RC
1 and the measured water hardness at the output RC 2.
In the programme steps 8.1 and 8.2 the desired measurement values are allo-
cated to the current values.
Further information on the recorder link can be found on page 9 in the section
"Recorder connection"
Connection and programming examples
Horn at the output OUT 1:
Function:
If the determined limit value is exceeded, the horn sounds for 60 seconds after every analysis.
Programming:
1.5 = MAX
= Activation of the horn if the water hardness is exceeded (applies to all outputs)
5.1 = IMP
= Generation of a pulse signal
5.4 = IMP | = Activation of the horn function by operation of the relay
6.1 = 60s
= Horn activation of 60 seconds duration
6.2 = 1 *
= Activation of the horn when the limit value is exceeded for the first time
In addition, the indicator type, the water hardness unit, a correction factor and the water hardness limit must
be specified in programme steps 1.1 - 1.4.
C
N
1
OUT 1
OUT 2
Power in
Protective earth
No
Nc
C
No
Nc
OUT 3
C
No
Nc
IN 1
IN 2
RC 1
RC 2
+
+
L
1
Power out
n
1
l
1
PE
2 0
2 1
1 9
1 8
1 7
1 6
1 5
1 4
1 3
1 2
1 1
1 0
9
8
2 4
2 3
2 5
2 6
1
3
4
5
6
7
22
PE
2
N
L
Ho
rn
So
le
n
o
id
v
a
lv
e
(op
e
n
ing
w
h
en l
iv
e
)
F
a
u
lt
al
arm
sy
ste
m
F
lo
w
m
o
n
ito
r
De
le
te
r
e
la
y
s
Re
c
o
rd
e
r 1
Re
c
o
rd
e
r 2
Po
w
e
r
c
o
n
nec
ti
o
n
Example RC1
18
+
19
RC 1
Example RC2
20
+
21
RC 2