Messages
6
SYCON
3000 C
Messages
If the integrated buzzer is activated in the course of a message, you can delete it by pressing the "HORN"
key. The message in the LCD display is only removed when the cause of the message is remedied or a new
analysis is started manually.
You acknowledge an activated relay by pressing the "RELAY" key.
In the case of the display "Analysis stop", you must start a new analysis manually. See the section "Manual
control" on page 5.
Refill indicator
I n d i c a t o r
r e f i l l
P o s s i b l e a n a l y s e s 2 4
H O R N
Indicator must be refilled. On the lower right-
hand side the probable number of possible analy-
ses is displayed.
As soon as this message is displayed, a certain amount of in-
dicator is available. As the indicator required per analysis is
dependent on the water hardness, the probable number of
analyses cannot be determined exactly in advance. In calculat-
ing the display value, soft water with a water hardness of
0 mg/l is assumed.
After each analysis this value is recaluculated with the actual
amount of indicator that still remains.
If there is no indicator when the instrument is
switched on, the display "Indicator lack Analysis
stop" appears immediately.
Indicator lack analysis stop
A T T E N T I O N !
I n d i c a t o r l a c k
A
n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Refill indicator.
Error blank value
A T T E N T I O N !
E r r o r b l a n k v a l u e
A
n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Possible causes:
Instrument or sample contaminated
No flush occurred
No water inflow
Calibration
necessary
Electrical defect (check connector)
Internal measurement too high
A T T E N T I O N !
I n t r . m e a s u r . t o o h i g h
A
n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Possible causes:
Indicator has not been metered out
No water inflow
Calibration
necessary
Electrical defect (check connector)
Internal measurement too low
A T T E N T I O N !
I n t r . m e a s u r e . t o o l o w
A n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Possible causes:
Mögliche Ursachen:
No water inflow
Calibration
necessary
Electrical defect (check connector)
Water hardness over/underrun
A T T E N T I O N !
W a t e r h a r d n e s s
e x c e e d e d
H O R N
According to the programming in program step
1.5, the overrunning or underrunning of the water
hardness is displayed.
Example: Hard water = Limit value exceeded.
Permanent signal 1
A T T E N T I O N !
P e r m a n e n t s i g n a l 1
A n a l y s i s s t o p
R E L A Y
Following an overrun or underrun of the water
hardness (see programming step 1.5), the activa-
tion of the relay "Permanent signal 1" is displayed.
It is also displayed whether an analysis stop has
occurred or whether analyses continue to be con-
ducted (see program step 6.3).
Permanent signal 2
A T T E N T I O N !
P e r m a n e n t s i g n a l 2
A n a l y s i s s t o p
R E L A Y
Following an overrun or underrun of the water
hardness (see program step 1.5), the activation of
the relay "Permanent signal 2" is displayed. This
relay can be activated even in the case of a fault
according to the programming in program step
6.7. It is also displayed whether an analysis stop
has occurred or whether analyses continue to be
conducted (see program step 6.3).