GX-2012 Operator’s Manual
Using Maintenance Mode • 119
Updating the Calibration Reminder Setting
With
CL RMNDR
set to
ON
(factory setting), the GX-2012 will give an indication at
start up if it is due for calibration. The type of indication will depend on the
CL EXPRD
setting (see below).
1. Use the AIR
or (SHIFT)
buttons to display the
CL RMNDR
menu item. Press
and release the POWER ENTER button. The screen will indicate the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
Updating the Calibration Expired Action Setting
This item defines what indication is given during start up when calibration is due and
CL
RMNDR
is set to
ON
.
With
CL EXPRD
set to
CONFIRM
(factory setting), the GX-2012 will give an
indication at start up if calibration is due and require the user to press the RESET
SILENCE button to continue.
With
CL EXPRD
set to
NOT USE
, if the unit is due for calibration the GX-2012 will
give an indication at start up that calibration is due and the unit cannot be used until it is
calibrated. With this setting, it is necessary to calibrate the instrument from Maintenance
Mode if calibration is not performed before it is due. See “Calibrating Using Auto
Calibration” on page 97 and “Calibrating Using Single Calibration” on page 100 for
instructions.
With
CL EXPRD
set to
NO EFFECT,
no confirmation during startup or calibration is
required to use the GX-2012.
1. Use the AIR
or (SHIFT)
buttons to display the
CL EXPRD
menu item. Press
and release the POWER ENTER button. The screen will indicate the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting along the bottom of
the screen.
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
On
MAINTENANCE
CL RMNDR
CAL
SEL
MAINTENANCE
CONFIRM