GX-2012 Operator’s Manual
Using Maintenance Mode • 127
2.
Press and release the POWER ENTER button. The screen will indicate the current
setting.
3.
Use the AIR
or (SHIFT)
button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and return to the main
menu.
Updating the Leak Check Mode CO Channel Display Setting
With
CO DISP
set to
OFF
(factory setting), the CO channel is not displayed in Leak
Check Mode but is still displayed in Normal Mode.
With
CO DISP
set to
On,
the CO channel will be displayed in Leak Check Mode but
there will be no CO alarms and the CO channel reading will not be saved to any snap log
data taken during the operating session.
1.
Use the AIR
or (SHIFT)
buttons to display the
CO DISP
menu item.
2.
Press and release the POWER ENTER button. The screen will indicate the current
setting.
3.
Use the AIR
or (SHIFT)
button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and return to the main
menu.