70 • Start Up, Bar Hole Mode
GX-2012 Operator’s Manual
4. Use the AIR
or (SHIFT)
buttons to scroll to the Bar Hole Mode Select Screen.
5. With the Bar Hole Mode Select Screen displayed, press and release the POWER
ENTER button to begin the Bar Hole Mode startup sequence.
NOTE:
If no button is pressed for 20 seconds, the unit will proceed into whichever
mode is displayed.
6. If
CL RMNDR
is turned on (factory setting), the screen that appears next depends on
how
CL EXPRD
is set in the Maintenance Mode Menu (page 119).
•
If the unit is due for calibration and
CL EXPRD
is set to CONFIRM (factory
setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After this, press and
release the RESET SILENCE button to continue.
•
If the unit is due for calibration and
CL EXPRD
is set to NOT USE, then the
following screen displays.
The GX-2012 cannot be used until a calibration has been performed either by
selecting
AUTO CAL
or
ONE CAL
in the Maintenance Mode menu. See
“Calibrating Using Auto Calibration” on page 97 or “Calibrating Using Single
Calibration” on page 100 for calibration instructions.
0
20.9
30
vol%
%LEL
ppm
%
BAR HOLE MODE
vol%
sec
CH
4
O
2
B.H. MODE?
CAL
C--LIMIT
FAIL
C--LIMIT