GX-2012 Operator’s Manual
Measuring Mode, Alarms • 41
Responding to Battery Alarms
WARNING:
The GX-2012 is not operational as a gas monitoring device during a
dead battery alarm. Take the GX-2012 to a non-hazardous area and
replace or recharge the batteries as described in “Replacing or
Recharging the Batteries” on page 134.
The GX-2012 is fully functional during a low battery warning. However, only a limited
amount of operating time remains, approximately 1 - 2 hours. The amount of time depends
on how often the LCD backlight is used and how often the unit is responding to alarm
conditions. Recharge the battery pack or replace the alkaline batteries as soon as possible
as described in “Replacing or Recharging the Batteries” on page 134.
NOTE:
Alarms and the back light feature consume battery power and reduce the
amount of operating time remaining.
Responding to Sensor Failure Alarms
1. Determine which sensor has triggered the sensor failure alarm.
2. Try calibrating the sensor first, as described in “Calibration Mode” on page 56 before
replacing it.
3. If the sensor failure continues, replace the sensor as described in “Replacing a
4. If the sensor failure condition continues after you have replaced the sensor, contact
RKI Instruments, Inc. for further instructions.
Responding to Clock Failure Alarm
This alarm occurs if the internal unit date has been changed to something unreasonable
like 15/34 (month/day).
1. Press and release the RESET SILENCE button to continue into normal operation.
CAUTION:
There will be no datalogging function if you operate the instrument after a
clock failure.
2. Attempt to change the date using the
DATE
menu time in Maintenance Mode. See
“Updating the Date and Time Settings” on page 95.
3. If the date cannot be set correctly, contact RKI Instruments, Inc. as soon as possible.
Responding to System Failure Alarms
1. If a system failure occurs, the system failure screen will display an error code as
shown below:
FAIL
031
SYSTEM
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30