118 • Using Maintenance Mode
GX-2012 Operator’s Manual
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
Updating the Datalog Overwrite Setting
With
OVER WRT
set to
ON
(factory setting), the GX-2012 writes over the oldest data
with new data when the data logger memory is full.
With
OVER WRT
set to
OFF
, the GX-2012 stops saving data to the data logger when the
data logger memory is full.
1. Use the AIR
or (SHIFT)
buttons to display the
OVER WRT
menu item. Press
and release the POWER ENTER button. The screen will indicate the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
Updating the Calibration Interval
This
CL INTVL
setting defines the amount of time between calibrations. The minimum
setting is 1 day and the maximum setting is 365 days. The factory setting is 90 days.
1. Use the AIR
or (SHIFT)
buttons to display the
CL INTVL
menu item. Press and
release the POWER ENTER button. The screen will indicate the current setting.
2. Use the AIR
or (SHIFT)
button to display the desired setting.
Tip:
Press and hold the
AIR
or
(SHIFT)
button to rapidly scroll
through the settings.
3. Press and release the POWER ENTER button to save the setting and return to the
main menu.
On
MAINTENANCE
OVER WRT
90
MAINTENANCE
CL INTVL