62
Serial Interface
DOD?
Get measurement data
Returns a value corresponding to the display indication.
Output data format
p1 p2 p3 p4 E ± p5, p6 <
EOT
><
CR
><
LF
>
p1 to p4
Effective 4 digits of measurement value
p5=-6 to +1
p6=O (overload
occurred)
p6= _
(no overload occurred) (_ is a space)
* Explanation of p5
p5 stands for 10
p5
.
For example, when measurement value is 10000,
p5 will be 1, because 1000 × 101
* When operating in maximum hold mode, returned
value corresponds to MAX HOLD value.
Same applies to peak hold mode.
When this command is received while there is no
display indication, output will be p1 to p4=0000.