19
Preparations
4. Use the setting keys ( , ) to set the display indication to the value
indicated as charge sensitivity on the calibration chart of the accel-
erometer.
When an accelerometer with integrated preamplifi er is used, enter
the voltage sensitivity. For example, when the voltage sensitivity is
5.90 mV/(m/s
2
), use the value 5.90 pC/(m/s
2
).
For the servo accelerometer LS-10C or LS-20C, enter the value ac-
cording to the calibration data (approx. 0.300).
When using the preamplifi er VP-26A, set the charge sensitivity.
Holding one of the setting keys down for more than 2 seconds causes
the display value to change quickly.