36
Measurement
Peak hold
The peak hold function uses 51.2 kHz sampling on the vibration input signal
to determine the peak value.
Setting procedure
1. Use the MENU key to call up menu 3.
2. Use the setting keys to select the setting 3-1 (peak hold ON).
3. Return to the measurement screen.
4. When the peak hold function has been activated, the indication PEAK HOLD appears
on the display, and the indicated numeric value is the peak value measured up to that
point. When a higher peak occurs, the display is updated.
When the peak hold value is higher than the overload threshold, the indication PEAK
HOLD OVER is shown.
5. The HOLD RESET key can be used to reset the peak hold value at
any time.