Chapter 2 To Set the Vertical System
RIGOL
MSO5000 User's Guide
2-3
To Adjust the Vertical Scale
Vertical scale indicates the voltage value per grid in the vertical axis of the screen. It
is often expressed in V/div. While you adjust the vertical scale, the display amplitude
of the waveform would enlarge or reduce. The scale information of the channel
status label (e.g. as shown in the following figure) at the lower section of the screen
would change accordingly.
The adjustable range of the vertical scale is related to the currently set probe ratio
and input impedance. By default, the probe ratio is 1X and the input impedance is 1
MΩ. In this case, the adjustable range of the vertical scale is from 1 mV/div to 10
V/div.
When CH1 is turned on, you can adjust the vertical scale with the following three
methods:
Rotate
Vertical
SCALE
that corresponds to CH1 to adjust the vertical scale
(clockwise to reduce the scale and counterclockwise to increase).
Enable the touch screen function and adjust the vertical scale with the pinch &
stretch gesture on the touch screen. For details, refer to the
section.
Enable the touch screen function, and tap the channel status label at the bottom
of the screen. The following window is displayed. Tap the
icon at the
right side of the
Scale
input field to decrease or increase the scale. You can also
tap the
Scale
input field to input a specific value with the displayed numeric
keypad.
In the vertical control area (Vertical) on the front panel, press
Vertical
SCALE
to adjust the vertical scale in "Coarse" (by default)
or "Fine" mode. You can also
press
1
More
Fine
to enable or disable fine adjustment.
Fine adjustment: Rotate
Vertical
SCALE
to further adjust the vertical scale
within a relatively smaller range to improve vertical resolution. If the amplitude
of the input waveform is a little bit greater than the full scale under the current
scale and the amplitude would be a little bit lower if the next scale is used, fine
adjustment can be used to improve the amplitude of waveform display to view
signal details.
Coarse adjustment (take counterclockwise as an example): Rotate
Vertical
Vertical Scale
Value Input Field Decrease Increase
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Summary of Contents for MSO5000 Series
Page 2: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 52: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 64: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 70: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 82: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 184: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 190: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 196: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 256: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 262: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 268: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 274: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 280: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 356: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...