RIGOL
Chapter 17 Function/Arbitrary Waveform Generator (Option)
17-2
MSO5000 User's Guide
To Output Basic Waveforms
Press
GI
on the front panel or enable the touch screen to tap the GI label at the
bottom of the screen to enable/disable the output of the
[GI]
connector on the front
panel, and enter the Function/Arbitrary Waveform Generator setting interface.
MSO5000 series' built-in Function/Arbitrary Waveform Generator can output a variety
of basic waveforms, including Sine, Square, Ramp, Pulse, DC, Noise, Sinc, Exp.Rise,
Exp.Fall, ECG, Gauss, Lorentz, Haversine, and Arb.
To Output Sine
Press
Wave
, rotate the multifunction knob
to select "Sine". Press down the
knob to select it. You can also press
Wave
continuously or enable the touch screen
to select it. At this time, you can set the parameters for the sine waveform.
1.
Set the frequency or period
First select
Frequency
or
Period
, then rotate the multifunction knob
to
select "Frequency" or "Period". You can also press
Frequency
or
Period
continuously to select it. Also, you can enable the touch screen to select it.
Rotate the multifunction knob
or use the pop-up numeric keypad to set the
frequency or period value of the current signal. Different waveforms have
different frequency or period (reciprocal of the frequency) ranges.
Sine: 100 mHz to 25 MHz
Square: 100 mHz to 15 MHz
Ramp: 100 mHz to 1 MHz
Pulse: 100 mHz to 10 MHz
DC and Noise: no frequency parameter.
2.
Set the amplitude or high level
First select
Amplitude
or
High Level
, then rotate the multifunction knob
to
select "Amplitude" or "High Level". You can also press
Amplitude
or
High
Level
continuously to select it. Also, you can enable the touch screen to select it.
Rotate the multifunction knob
or use the pop-up numeric keypad to set the
amplitude or high level value of the current signal. Note that if you select
"Amplitude" for this menu, then the menu below the
Amplitude
is automatically
turns out to be "Offset"; if you select "High Level" for this menu, then the menu
below the
High Level
is automatically turns out to be "Low Level". When
Impedance is set to "HighZ", the available range is from 20 mVpp to 5 Vpp; when
Impedance is set to "50Ω", the available range is from 10 mVpp to 2.5 Vpp.
3.
Set the offset or low level
First select
Offset
or
Low Level
, then rotate the multifunction knob
to
select "Offset" or "Low Level". You can also press
Offset
or
Low Level
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Summary of Contents for MSO5000 Series
Page 2: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 52: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 64: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 70: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 82: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 184: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 190: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 196: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 256: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 262: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 268: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 274: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 280: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Page 356: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...