RIGOL
DP1116A Performance Verification Manual
1-3
Voltage and Current Values
During the test, you need to set the voltage and current of the output scale of the
power supply to specified values. Table 1-1 lists the rated output values and
maximum output values of the voltage and current of the two scales.
Table 1-1 Voltage and Current Values of the Two Scales of DP1116A
Scale
Rated Output
Voltage
Max Output
Voltage
Rated Output
Current
Max Output
Current
+32V/5A
+32 V
+33.6 V
5 A
5.25 A
+16V/10A +16 V
+16.8 V
10 A
10.5 A