DP1116A Performance Verification Manual
Test Preparations
You need to make the following preparations before performing the performance
verification test.
1.
Perform self-test to make sure that the instrument can work normally.
Connect the instrument to AC power supply using the power cord provided with
the accessories. Turn on the power switch at the rear panel; then press the
power key
at the front panel; the instrument starts and performs self-test.
If the instrument passes the self-test, the user interface is displayed; otherwise,
the self-test error prompt message is displayed. You can press Utility
SelfTest
Test to perform self-test after the instrument starts. If
the self-test fails, make sure that the problems are found and solved and the
instrument passes the self-test before performing the performance verification
test.
Note:
Before connecting the power, make sure that the setting of the voltage
selector at the rear panel of the power supply matches the AC power supply
to be connected. DP1116A supports the following input voltages.
100: 100 Vac
±
10%, 50 Hz to 60 Hz, T4 A fuse
115: 115 Vac
±
10%, 50 Hz to 60 Hz, T4 A fuse
220: 220 Vac
±
10%, 50 Hz to 60 Hz, T2.5 A fuse
230: 230 Vac
±
10%, 50 Hz to 60 Hz, T2.5 A fuse
Before performing self-test, make sure that there is no connection at the
front and rear panel terminals of the power supply.
2.
Make sure that the power supply is within the calibration period (1 year). If
calibration is required, please contact
RIGOL
.
3.
Run the instrument for at least 30 minutes.
4.
Make sure that the environment temperature is between
20℃
and
30℃
and the
relative humidity is less than 80
%
.
5.
The test connecting wires used should be as short as possible. Before
performing the test, press Sense to enable the Sense mode.
Note:
The test introduced in this manual should be done by professionals. During the test,
there might be dangerous voltage at the output terminal of the power supply and
some test devices.