Chapter 2 Constant Voltage Test
RIGOL
DP1116A Performance Verification Manual
2-3
Explanation:
The short-circuit devices in Figure 2-1 are used to short-circuit the (S+) terminal
and (+) terminal as well as the (S-) terminal and (-) terminal respectively. Before
performing the test, press Sense to enable the Sense mode. The performance test
measures the specifications under Sense mode (with short-circuit devices).