Chapter 2 Constant Voltage Test
2-14
DP1116A Performance Verification Manual
Test Record Form:
Table 2-5 CV Programming and Readback Accuracy Test Record Form
Scale
+16V/10A
+32V/5A
Specification
Programming:
0.05%*Volt+ 10 mV=18
mV
Readback: 0.05%*Volt+
5 mV=13 mV
Programming:
0.05%*Volt+ 10 mV=26
mV
Readback: 0.05%*Volt+ 5
mV=21 mV
Output voltage is 0
DP1116A
Setting
Volt (V)
0
0
Current (A)
10.5
5.25
Measurement
Value
Test Result
U
1
U
2
Programming
Accuracy: |U
1
-0|
Readback
Accuracy: |U
2
-0|
Output voltage is the rated output voltage
DP1116A
Setting
Volt (V)
16
32
Current (A)
10.5
5.25
Measurement
Value
Test Result
U
3
U
4
Programming
Accuracy:
|U
3
-rated output
voltage of the
scale under test|
Readback
Accuracy:
|U
4
-rated output
voltage of the
scale under test|