RIGOL
DP1116A Performance Verification Manual
3
CV Programming and Readback Accuracy Test Record
Form
Scale
+16V/10A
+32V/5A
Specification
Programming:
0.05%*Volt+ 10 mV=18
mV
Readback: 0.05%*Volt+
5 mV=13 mV
Programming:
0.05%*Volt+ 10 mV=26
mV
Readback: 0.05%*Volt+ 5
mV=21 mV
Output voltage is 0
DP1116A
Setting
Volt (V)
0
0
Current (A)
10.5
5.25
Measurement
Value
Test Result
U
1
U
2
Programming
Accuracy: |U1-0|
Readback
Accuracy: |U
2
-0|
Output voltage is the rated output value
DP1116A
Setting
Volt (V)
16
32
Current (A)
10.5
5.25
Measurement
Value
Test Result
U
3
U
4
Programming
Accuracy:
|U
3
-rated output
voltage of the
scale under test|
Readback
Accuracy:
|U
4
-rated output
voltage of the
scale under test|
CC Load Regulation Test Record Form
Scale
16V/10A
32V/5A
Specification
<0.005%*250
μA =750 μA
<0.005%*250
μA=500 μA
DP1116A
Setting
Volt (V)
16.8
33.6
Current (A)
10
5
Measurement
Value
U
0
U
1
|U
1
/R - U
0
/R|