User’s Guide for DG2000 Series
2-82
1.
View the last message
View the last message
2.
View the remote Command error queue
View the remote command error queue in the remote operation.
3.
Get HELP on any key
To get context-sensitive help on any front-panel key or menu softkey, press and
hold that key for a few times.
4.
Generate a DC-only Voltage level
(1)
Press Utility and select DC On .
(2)
Press Offset, then enter the desired voltage level.
(3)
Choose DC On/Off
Æ
DC Off , or press any function key to resume the
normal waveform operation.
5.
Generate a modulated Waveform
(1)
Configure the Carrier waveform by selecting the waveform shapes, frequency,
amplitude, etc. Pulse can only be modulated using PWM; Noise and DC can
not be modulated.
(2)
Configure the modulating waveform by pressing the Mod key.
(3)
To change the Carrier parameter, press the lighted function key (Sine,
Square etc).
(4)
To turn off the modulation, press the lighted Mod key.
6.
Create an arbitrary waveform
(1)
Press Arb and choose “ Create New ”.
(2)
Enter the desired period, voltage limit, and the number of points and then
press “ Edit Points ”, the first two points are defined for you.
(3)
Rotate the knob to select the desired point for editing and then use Time
and Voltage to define the point. Press Insert Point to insert additional
points as needed.
(4)
The waveform editor connects the last point to the voltage level of point #1
to create a continuous waveform. Make sure that the time value of the last
point less than the specified period.
(5)
Pres End/Store to store the waveform and exit the waveform editor.
7.
Reset instrument to its default state