RIGOL
© 2006 RIGOL Technologies, Inc
User’s Guide for DG2000 Series
4-8
has to be turned off to allow Pulse.
Sweep has been turned off to allow Noise.
Pulse, noise and DC are not allowed to generate Sweep waveform. So the modulation
has to be turned off to allow Noise.
Sweep has been turned off to allow DC.
Pulse, noise and DC are not allowed to generate Sweep waveform. So the modulation
has to be turned off to allow DC.
Manual Trigger only for Sweep, N-Cycle Burst.
Manual Trigger is only used for Sweep and N-Cycle Burst.
Trigger Source has been changed to Manual.
When Trigger is enabled, the source type turns from internal to manual.
Pulse width was reduced by period.
For a pulse waveform, the generator will adjust the waveform parameter
automatically to fit the need of the pulse according to the order below: Edge time-
Pulse Width-Period. In this case, the generator will reduce the pulse width to cater to
the need of the period. The Edge Time is already set to the minimum value.
Edge Time was reduced by period.
For a pulse waveform, the generator will adjust the waveform parameter
automatically to fit the need of the pulse according to the order below: Edge time-
Pulse Width-Period. In this case, the generator will reduce the Edge Time to cater to
the need of the period.
Amplitude (V
rms
) upper limit =3.536V
rms
For example: if a square wave with the amplitude of 5 V
rms
connects to a 50 ohm
resistance was turned into a sine waveform, then the generator will change the
amplitude to 3.536 V
rms
automatically.
Offset has changed due to Amplitude.
In DC voltage function, voltage level is adjusted through the change of the offset, and
the current amplitude will be ignored. When changing to the other function, the
generator will adjust the offset voltage to fit the current amplitude setting.
Deviation has been changed due to carrier frequency.
The Carrier frequency should always be greater than or equal to the deviation
frequency. If the carrier frequency is set less than the deviation frequency, when FM is
activated, the generator will adjust the deviation frequency to the value automatically
that the current carrier frequency allows.
Duty cycle is limited at a higher frequency.
If the waveform is set Square and the frequency is set to a value which does not allow
the current Duty Cycle, the Duty Cycle will adjust to the value automatically that the
current frequency allows. For example, if the current Duty Cycle is 70%, and the