RIGOL
© 2006 RIGOL Technologies, Inc
User’s Guide for DG2000 Series
4-15
Square or Sine wave, the Center Frequency upper limit is 34MHz; If the Carrier is a
Ramp wave, the Center Frequency upper limit is 400kHz; If the Carrier is an Arb wave,
the Center Frequency upper limit is 12MHz.
Center Frequency lower limit = 1
μ
Hz
The Center Frequency lower limit is 1
μ
Hz. If the specified Center Frequency is lower
than 1
μ
Hz, the system will adjust the Center Frequency to 1
μ
Hz automatically.
Frequency span upper limit = **
With different Source type modes and center frequencies, the frequency span upper
limits vary.
Frequency span lower limit = **
With different Source type modes and center frequencies, the frequency span lower
limits vary.
Sweep Time upper limit = 500.00s
The Sweep Time upper limit is 500.00s. If the specified Sweep Time is greater than
500.00s, the system will adjust the Sweep Time to 500.00s automatically.
Sweep Time lower limit = 1.0ms
The Sweep Time lower limit is 1.0ms. If the specified Sweep Time is less than 1.0ms,
the system will adjust the Sweep Time to 1.0ms automatically.
Phase upper limit= +360.0º
The Phase upper limit is +360.0
o
. If the specified Phase is greater than +360.0
o
, the
system will adjust the Phase to +360.0
o
automatically.
Phase lower negative limit
=
- 360.0º
The Phase lower negative limit is -360.0
o
. If the specified Phase is less than -360.0
o
,
the system will adjust the Phase to -360.0
o
automatically.
# of Cycles upper limit = 1,000,000 Cyc
The Cycle upper limit is 500,000Cyc. If the specified Cycle is greater than 500,000Cyc,
the system will adjust the Cycle to 500,000Cyc automatically.
# of Cycles lower limit = 1Cyc
The Cycle lower limit is 1Cyc. If the specified Cycle is less than 1Cyc, the system will
adjust the Cycle to 1Cyc automatically.
Trigger Period upper limit = 500.00s
The Trigger Period upper limit is 500.00s. If the specified Trigger Period is greater than
500.00s, the system will adjust the Trigger Period to 500.00s automatically.
Trigger Period lower limit = 62.5
μ
s
The Trigger Period upper limit is 62.5
μ
s. If the specified Trigger Period is less than
1.0
μ
s, the system will adjust the Trigger Period to 62.5
μ
s automatically.
Delay upper limit = 40s
The Delay upper limit is 40s. If the specified Delay is greater than 40s, the system will