User’s Guide for DG2000 Series
3-16
Example 11: To generate a Linear Sweep
How to generate a Sine Sweep Waveform whose frequency starts from 100Hz to
10kHz. Use the Internal Trigger Mode, Linear Sweep, and the Sweep time is 1s.
The operation steps:
1.
Choose the Sweep Function:
Press Sine, and choose the Sine Waveform as the Sweep Function.
The default setting of the Source is Internal.
2.
Set the Frequency, Amplitude and Offset.
(1)
Press Period/Freq and choose Freq which will display in reverse color.
Enter “5” from the keypad and choose the unit “kHz” to set the Frequency as
5 kHz.
(2)
Press Ampl/HiLev and choose Ampl which will display in reverse color.
Enter “2” from the keypad and choose the unit “V
pp
” to set the Amplitude as
2 V
pp
.
(3)
Press Offset/LoLev and choose Offset which will be displayed in
reverse color. Enter “0” from the keypad and choose the unit “V
dc
” to set the
Offset as 0 V
dc
.
You can view the Sweep Waveform in the Graph Mode.
3.
Set the Sweep Mode.
Press Mod
Æ
Linear/Log and choose Linear. Please notice that the message
shown on the top left side of the screen is “Linear”.
4.
Set the Sweep Time
Press Time , use the keypad to enter “1”, choose the unit “s”, to set the Sweep
Time as 1s.
5.
Set the Start Frequency
Press Start , use the keypad to enter “100”, choose the unit “Hz”, to set the Start
Frequency as 100Hz.
6.
Set the Stop Frequency
Press Stop , use the keypad to enter “10”, choose the unit “kHz”, to set the Stop
Frequency as 10 kHz.
At this time, the instrument generates the specified Sweep waveform whose