![Rigol DG2000 Series User Manual Download Page 142](http://html1.mh-extra.com/html/rigol/dg2000-series/dg2000-series_user-manual_1461078142.webp)
RIGOL
© 2006 RIGOL Technologies, Inc
User’s Guide for DG2000 Series
4-12
low level, the generator will set the low level 1 mV less than the high level
automatically.
Duty Cycle upper/lower limit = **
In the Square mode, different frequency range corresponds to different Duty Cycle
limit.
For a frequency lower than 25MHz, the range is from 20% to 80%;
For a frequency between 25 ~ 50MHz, the range is from 40% to 60%;
For a frequency higher than 50MHz, the Duty Cycle is set to be 50%.
Symmetry upper limit = 100.0%.
In the Ramp mode, the Symmetry upper limit is 100%. If the specified symmetry is
greater than 100%, the system will adjust the symmetry to 100% automatically.
Symmetry lower negative limit = 0.0%.
In the Ramp mode, the Symmetry lower limit is 0.0%. If the specified symmetry is
less than 0.0%, the system will adjust the symmetry to 0.0% automatically.
Pulse Width is limited by Period.
The Pulse Width should be less than the difference of the period and the Edge Time.
Pulse Width< period- 1.6×Edge Time.
Edge Time is limited by Period
The specified Edge Time should be in the specified period. The generator will adjust
the Edge Time to meet the need of the period. Edge Time < 0.625×Pulse Width.
Pulse width lower limit = 12ns
In the Pulse mode, the Pulse width lower limit is 12ns. If the specified pulse width is
less than 12ns, the system will adjust the pulse width to 12ns automatically.
Edge Time lower limit = 5ns.
In the Pulse mode, the Edge Time lower limit is 5ns. If the specified Edge Time is less
than 5ns, the system will adjust the Edge Time to 5ns automatically.
Offset upper limit= **
Different mode has different Offset upper limit; Offset upper limit may vary in the
same mode if the load is different.
Offset lower negative limit = **
Different mode has different Offset lower limit; Offset lower limit may vary in the same
mode if the load is different.
Initial # of points upper limit = 524,288
The number of initial points upper limit is 524
,
288. If the specified waveform has more
than 524
,
288points, the system will adjust the point number to 524
,
288 automatically.
Initial # of points lower limit=2.
The default initial point number is 2, the lower limit of the initial points. If the specified
initial point is less than 2, the system will adjust the number to 2 automatically.