Guidelines and manufacturer's declaration – electromagnetic immunity
The ri-scope® L/EliteVue instruments are intended for use in the electromagnetic environ-
ment specified below. The customer or user of the ri-scope® L/EliteVue should ensure that it
is used in such an environment.
Immunity testing
IEC 60601
test level
Compliance
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Con: ±8 kV
Air: ±2, 4, 8,
15 kV
Con: ±8 kV
Air: ±2, 4, 8,
15 kV
Floors should be wood, concrete or
ceramic tile.
If the floor is covered with synthetic
material, the relative humidity must be
at least 30%.
Fast transient
electrical distur-
bances / bursts
IEC 61000-4-4
5/50 ns, 100
kHz, ±2 kV
Not applicable
The quality of the supply voltage should
be that of a typical commercial or
hospital environment.
Impulse voltage
IEC 61000-4-5
±0.5 kV voltage
Phase-to-pha-
se conductor
± 2 kV voltage
Line-to-earth
± 0.5 kV
voltage
Phase-to-pha-
se conductor
± 2 kV voltage
Outer conduc-
tor to earth
Not applicable
The quality of the supply voltage should
be that of a typical commercial or
hospital environment.
IEC 61000-4-11
Voltage
dips, short
interruptions and
voltage variations
according to
IEC 61000-1-11
<0% UT 0.5
period at
0, 45, 90, 135,
180, 225,
270 and 315
degrees
0% UT 25/30
periods
Single phase:
at 0 degrees
(50/60 Hz)
Not applica-
bled
The quality of the supply voltage
should be that of a typical commerci-
al or hospital environment.
Magnetic field
with energy-ef-
ficient rated
frequencies
IEC 61000-4-8
30A/m
50/60 Hz
30A/m
50/60 Hz
Mains frequency magnetic fields
should be at a level characteristic of
a typical location in a typical com-
mercial hospital environment.
Note: U
T
is the AC source. Mains voltage before the application of the test level.
33