20
DC-100 Operation Manual
604
Weight Decimal Point Position for Scale A
0: 0
1: 0.0
2: 0.00
3: 0.000 (DEFAULT)
4: 0.0000
605
Load Cell Sensitivity Selection in mV/V for Scale A
0: 4.00 mV/V
1: 3.76 mV/V
2: 3.52 mV/V
3: 3.28 mV/V
4: 3.04 mV/V
5: 2.80 mV/V
6: 2.56 mV/V
7: 2.32 mV/V
8: 2.08 mV/V
9: 1.84 mV/V
10: 1.60 mV/V
11: 1.36 mV/V
12: 1.12 mV/V (DEFAULT)
13: 0.88 mV/V
14: 0.64 mV/V
15: 0.40 mV/V
606
Load Cell Type for Scale A
0: Standard / Normal load cell (DEFAULT)
1: Abnormal load cell with large offset
607
A/D Board Type for Scale A
0: Normal (DEFAULT)
1: Prevent from small vibration / fast change in display
2: Prevent from medium vibration
3: Prevent from large vibration / slow change in display
608
External Scale B Connection
0: Disable (DEFAULT)
1: Enable
609
Minimum Display (Increment) for Scale B
0: 2 (DEFAULT)
1: 1
2: 5
3: 10
610
Weight Decimal Point Position for Scale B
0: 0
1: 0.0
2: 0.00
3: 0.000
4: 0.0000 (DEFAULT)
Spec
Description
Table 3-2. DC-100 (142) Settings