Calibration
55
4.0
Calibration
The
Counterpart
can be calibrated using the front panel, EDP commands, or Revolution
®
.
Note
Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 4-1 and the
SCALE sub-menu shown in Figure 3-8 on page 28. The CALIBR menu shown in Figure 4-1 is a “quick access”
calibration; for more in-depth scale setup and calibration, use the menus found under SETUP
/
CONFIG
/
SCALE (see Figure 3-8 on page 28).
Calibration consists of the following steps:
•
Zero calibration
•
Entering the test weight value
•
Span calibration
•
Optional five-point linearization
•
Optional rezero calibration for test weights using hooks or chains.
Note
The
Counterpart
requires the WZERO and WSPAN points to be calibrated. The linearity points are optional,
but must not duplicate zero or span. During calibration, the TARE (Enter) key acts as a data entry confirmation
key. The soft calibrate key accepts that value.
Display and edit
test weight value
Same as POINT 1
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Previous A/D raw
counts are shown
Press calibrate
softkey to calibrate
zero. Press menu
to cancel
calibrating.
Calibrating, please wait
Press enter to
remove offset from
zero and span
calibrations
POINT 1
POINT 2
POINT 3
POINT 4
POINT 5
WZERO
WVAL
WSPAN
REZERO
WLIN
WGT 1
CAL1
ID
DISPLAY
TEST
SETUP
AUDIT
CALIBR
SETPTS
Figure 4-1. Calibration (CALIBR) Menu
4.1
Front Panel Calibration
1. Press the
MENU SETUP
button, and then press until the display reads
CALIBR
.
2. Press
TARE (Enter)
or to go to the scale to be calibrated. Press
TARE (Enter)
or to go to
WZERO
. Press
TARE (Enter)
again to display a value and press the
Calibrate
softkey
to Calibrate
WZERO.
Note
If you do not want to calibrate zero, press Menu to exit.
3. The indicator displays
CALIBRATING
while calibration is in progress. When complete, press
TARE (Enter)
or
to advanced to the next prompt (
WVAL
).
Note
To view the new AD count, repeat Step 3, but instead of pressing TARE (Enter) while viewing the value, press
MENU SETUP to exit.
4. With
WVAL
displayed, press
TARE (Enter)
or
to display or edit the stored calibration weight value. Press
TARE (Enter)
to store the
WVAL
value and advance to
WSPAN
.
5. With
WSPAN
displayed, press
TARE (Enter)
or to view the A/D count. Place test weights on the scale equal
to WVAL. Press the
Calibrate
softkey to calibrate WSPAN.