Configuration
35
Parameter
Choices
Description
COUNT
NEGCOUNT
SAMPLEQTY
INSFSMPL
LOTUPDT
SCLCHG
NEWITEM
XFRSampl
DISPACC
UNITWTUPDATE
UTWTBASE
DSP Mode
RemovESP
Selects the counting mode (enables/disables negative checkweighing; turns data parameters on/
off. See Figure 3-12.
TARE RND
Exact Round
When transferring tare to the second scale, tare value remains the exact weight or rounds to
scale resolution.
REGION
REGULA
REGWRD
DECFMT
TIME
DATE
Selects regional settings. See Figure 3-13.
UID
1
Sets the unit ID, a string of up to 6 ASCII characters, which can be set via serial port or keypad.
This will be used in place of the <UID> token in a print format. The default value is 1.
SINGLE ID
Yes
No
Allows entry of ID to not be stored. Next ID clears field and establishes new data.
CONSC#
CURVAL
RESVAL
Allows sequential numbering for print operations.
• CURVAL is the current/starting value
• RESVAL is the ending/reset value
The consecutive number value is incremented following each print operation that includes <CN>
in the ticket format. When the consecutive number is reset, it is reset to the RESVAL specified on
the parameter.
PASSWD
CALIBR
SETUP
TEST
T&D
ID
Creates a password to access the CALIBR, SETUP, TEST, T&D, and ID menus. Specify a
non-zero value to restrict access to all configuration menus. Passwords can be overridden by
loading new firmware or entering 999999.
Note
Overriding passwords will clear configuration and calibration settings. To
preserve settings (such as ID information), use Revolution software to upload
your data to a PC, and then download it back to the indicator after the
password override is performed.
SOFTKEYS
SK1-12
See Figure 3-14.
KEYLCK
MENU
ZERO
GRSNET
UNITS
PRINT
TARE
NUMBER
SAMPLE
UNIT WT
Disables the MENU, ZERO, GROSS/NET, UNITS, PRINT, TARE numeric, SAMPLE and UNIT
WEIGHT keys. Select
LOCK
to disable the key, and
UNLOCK
to enable the key.
STNDBY
0
Standby mode delay. Specifies the number of minutes the indicator must be inactive before
entering standby mode. Valid values are 0 (off) through 255 minutes. When in standby mode,
power is still supplied to the CPU and draws half of the current as when the display is powered.
The display is no longer updated and the backlight is off. Press any key to exit standby mode and
reactivate the display.The indicator enters standby mode if no key presses, serial
communications, or scale motion occur for the length of time specified in this parameter. Set to 0
to disable standby mode.
Note
If regulatory agency is set for NONE or INDUST, the standby will occur after
the specified number of minutes regardless of how much weight is on any of
the scales. If regulatory agency is set for NTEP, OIML, or Canada a standby
will only occur when the weight on the currently viewed scale is at center of zero.
Table 3-10. Feature Menu Parameters