Operating the Device
25
Fig. 16:
End of program
Once the “End of program“ message appears, the press is blocked for approx. 25 seconds
in order to give the cylinder time to return to the initial position.
Open the door, pull out the tool sliding tray and take the cover off.
Put on the pressing-out aid.
Fig. 17:
Put on the pressing-out aid
Push the tool sliding tray into the press again.
Close the door.
Select [Sample output].
Fig. 18:
Activate the sample output.
As long as the cylinder is not in its initial position, a red bar appears in the display. As soon
as the cylinder has been pulled back to the initial position, the red bar disappears.