Only manufacturer approved components, specified for the Leo
bed, should be used - if in doubt contact your local distributor,
importer or manufacturer.
▪
Never attempt to re-wire any components.
▪
Over time, the auxiliary emergency power supply may emit an
increased amount of flammable gas, creating a risk of
explosion/fire. Reha-Bed sp.z o.o. recommends replacing the
batteries every 4 years or earlier.
▪
Check if all electrical functions operate correctly.
▪
Check if all electrical cables are in good condition.
▪
Check if the mains cable and plugs are in good condition, if either is damaged it
must be replaced as a complete assembly, the plug or cable must never be re-
wired.
▪
Check if cover protecting the actuator plugs and the power cable plug in the
control box is fitted.
▪
Check if all nuts, screws and fasteners are tight and that none are missing or
incomplete.
▪
Check if all screws and knobs are present.
▪
Check if the backrest section is mechanically operational.
▪
Check if the leg section (including the knee break functions) work correctly.
▪
Check if all labels and stickers are present and legible.
▪
Raise and lower the side rails – check if they move smoothly.
▪
Check if the lock on the side rails automatically engages when raised.
▪
Check if the castors lock function works correctly and that when locked castors do
not move.
▪
Check if the bed’s frame does not show signs of excessive wear (in particular
whether there are no cracks near welds).
▪
Check the condition and voltage of the batteries for the emergency power system.
It is recommended to replace them when the battery condition or nominal voltage
is abnormal.
▪
If any gaps appear to be outside of specification remove the bed from use until the
dimension of the gap in question has been confirmed.
If in doubt about correct replacement of a component contact your local distributor or
seller. Refer to the parts list for part codes and assembly detail. Copies are available
from your local distributor or importer. Contact details can be found in section 1.1 of
the booklet.
12.2.
S
ERVICE LIFE
The service life of the Leo MED bed is 10 years*, with the exception of the mattresses.
On the basis that the bed and its associated accessories are serviced and maintained