19
Make sure that all packaging parts that secure movable sections
(such as cable ties, foils, tapes, etc.) have been removed before the
bed is put into service
–
otherwise, there is a risk of damaging the
bed’s frame.
6.3.
F
ITTING THE MEDICAL SIDE RAILS
Leo MED is delivered with fitted medical side rails. The following instructions may be
helpful when you need to replace the side rails or their components.
▪
Insert the side rail’s p
ivots into the mounting
bush in the mattress platform (as shown on the
right).
▪
The side rails should be placed evenly into both
mounting bushes.
Note 1
: Make sure that the side rails release
button is positioned near the foot end.
Note 2
: The medical side rails are available in the right and left version!
▪
Ensure that the latches at the bottom of the mounting bushes lock the side rail
in place.
▪
If there is any doubt about the assembly contact the provider
or manufacturer, incorrectly fitted side rails can lead to
death.
▪
Side rails must be installed on both sides of the bed (on the
side of the wall as well).
▪
With standard side rails, the maximum height of the mattress
is 150 mm.
▪
In case of damage to the rails (bending, breaking, cracking
etc.), they should be immediately replaced with new ones
due to the risk of an accident.