REGULA 7517A User Manual Download Page 9

9

Components

Quantity

Packing

Figure №

Marking

Test-object

1

bag

1b, 2

Sleeve gaskets of the combination 

magnetic scanner

2

bag

Polyethylene bag

1

box

Transportation packaging

1

Regula 7517A

serial #,

year of 

production

** — supplied optionally.

Configuration and Operation

Regula 7517A is a set of hardware components and software products. The device allows 

examining objects using different methods. 

Examination starts with magnetic copying (making a magnetogram = magnetic copy) of 

the  ferromagnetic  object  surface.  The  magnetic  copying  accessories  set 

(MCAS)

  contains 

various  kinds  of  magnetic  scanners  which  locally  magnetize  the  object,  excite  its  magnetic 

stray fields and copy them to the intermediate flexible magnetic carrier (magnetic tape) without 

removing the lacquer–and-paint coating. 

The eddy-current scanner

 (supplied optionally) is 

used for investigating internal stresses in ferromagnetic objects and for examining the surface 

of aluminum objects.

The  magnetogram  is  loaded  into  the 

USB device for magneto-optical visualization

 

(

MOV USB device

) where information about object’s surface layer is scanned. The MOV USB 

device combines two functional components: optic-mechanical block and block of electronics.

The optic-mechanical block (OMB) 

contains a magneto-optical visualizer which converts 

magnetic stray fields of a magnetic copy into an image and a special tape drive mechanism 

which moves the magnetic copy relative to the visualizer.

A  magnetic  copy  is  loaded  into  the  tape  drive  where  it  is  automatically  detected  and 

introduced  into  the  tape  drive  tract.  Then  it  is  scanned  step-by-step  by  a  sensor  designed 

on the basis of a magneto-optical crystal. Magneto-optical visualization of magnetograms of 

the examined object is based on Faraday magneto-optical effect which occurs in crystalline 

films of Bi-containing ferrite garnets.

Light distribution obtained as a result of visualization is received by the sensitive element 

of the video camera and is converted into a digital TV signal which is further processed and 

examined with the help of PC software.

Step-by-step input of images into a PC allows making automatic panoramic stitching of a 

digital image from several sub-shots.

The block of electronics 

manages OMB  resources (electric  drive,  inductor,  illuminator, 

video camera and indicators).

The

 PC 

operates the MOV USB device via special software. The PC displays newly obtained 

and stored examination data which are processed, examined and documented.

Peripherals 

(supplied optionally) extend examination possibilities of the device by applying 

additional methods and means of investigation.

The device operates together with the following peripherals:

 

z

USB device for optical input,

 

z

magnetic powder visualization device,

 

z

eddy-current probe,

 

z

electrochemical etching device.

Basic functions of NUCA (EYER) 

software

:

 

z

operating USB devices,

 

z

input and processing of images, 

 

z

automatic program correction and panoramic stitching of MOV frames,

 

z

image scaling and rotation,

 

z

comparison of images,

 

z

measurements of linear and angular sizes,

 

z

saving and printing of image files as illustrations to the photo charts.

Table 1 (continued)

1.1.4

Summary of Contents for 7517A

Page 1: ...WARE AND SOFTWARE SYSTEM FOR FORENSIC EXAMINATION OF FIREARMS AND AMMUNITION SERIAL NUMBERS REGULA 7517 MAGNETO OPTICAL DEVICE REGULA 7517A USER S GUIDE Certified Quality Management System Regula Ltd...

Page 2: ......

Page 3: ...sories Set MCAS 13 1 2 3 Personal Computer PC 14 1 2 4 Software 15 1 2 5 Spare Parts and Repair Kit SPRK 15 2 INTENDED USE 16 2 1 Device Intended Use 16 2 1 1 Safety measures and operating limitations...

Page 4: ...ssories set MOI magneto optical image MOV magneto optical visualization MOV USB device USB device for magneto optical visualization OMB optic mechanical block MT magnetic tape SPRK spare parts and rep...

Page 5: ...ining and get clearance to perform expert examination in the field of trace evidence z z be an OS Windows user z z study techniques of authenticity verification of firearms and ammunition serial numbe...

Page 6: ...laboratory conditions due to the fact that the device provides high quality results visual evidence and uses non destructive magneto optical methods In case any changes in serial numbers are detected...

Page 7: ...than 5 years Operation warranty period 12 months from the date the device is delivered to the customer Storage warranty period 12 months Dimensions In transportation packaging not more than 350 280 22...

Page 8: ...sories 1 box 1b Peripheral examination tools Regula 7516 body case Regula 7516 Software PC drivers and utilities CD folder CD 1 OS Windows XP CD 2 MS Office CD 3 Special software for managing USB devi...

Page 9: ...where it is automatically detected and introduced into the tape drive tract Then it is scanned step by step by a sensor designed on the basis of a magneto optical crystal Magneto optical visualizatio...

Page 10: ...g with fusible metals dot welding and pressing in fragments See paragraph 2 2 1 2 for the details of the device operational check up using the test object Test object specifications The test object se...

Page 11: ...r s name and or trade mark z z manufacturer s model identification z z device serial number in compliance with the numbering system of the manufacturer z z date of manufacturing year and month z z rat...

Page 12: ...te time of input of a magnetic copy of the number platform 15 sec OS WindowsXPorlater Format of image files BMP PCX GIF JPG PC monitor size 12 Approximate size of an image file of the number platform...

Page 13: ...he copy reflects relief and structural inhomogeneity of the material Permanent magnets of the combination magnetic scanner CMS concentrators serve as sources of magnetic stray fields excitation z z De...

Page 14: ...ced examination of metal structure of objects is the eddy current scanner supplied optionally Its specifications are given in the user s guide to eddy current magnetographing device Regula 7517B Magne...

Page 15: ...ion results Detailed description of NUCA EYER software is given in the corresponding user s guide Spare Parts and Repair Kit SPRK Function SPRK is intended for routine maintenance and repairs of the d...

Page 16: ...z Consider operating limitations for the components of the device in conformity with paragraph 2 2 2 2 1 2 1 1 Failure to observe operating limitations given in paragraphs 2 1 1 and restrictions on th...

Page 17: ...rnal power supply source The device can be powered by the internal autonomous PC battery In this case power supply cables are not connected z z Switch on the PC Start the OS and launch NUCA EYER softw...

Page 18: ...d it will move forward automatically z z Wait for approximately 5 seconds to load the potentially informative sector of the magnetic copy into MOV zone Start the stitching process by corresponding sof...

Page 19: ...mined number Measure geometric dimensions of the images of the digits and analyze their position The software allows measuring linear and angular image dimensions z z Analyze examination results Docum...

Page 20: ...Figure 5a to move the MT strip Figure 5 pos 3 through the slot beyond the edge of the blade Figure 5 pos 2 z z Take the MT by its edge Figure 5 pos 3 and pull it out for a required length relative to...

Page 21: ...netized MT Inserting MT strips into the cartridge and extracting them Figure 6 d c b 1 cartridge for storing MT 2 magnetic clamp 3 MT 1 2 To insert MT strips into the cartridge do the following Roll u...

Page 22: ...irty MT surface greasy film water solvents chemical solutions particles of dust rust metal etc MT surface should be cleaned with alcohol using a special container from the SPRK Figure 7d MT with consi...

Page 23: ...tic MT in the conditions which improve the quality of magnetic copying e g for examining the relief and structure of aluminum objects for examining weak internal stresses in ferromagnetic objects for...

Page 24: ...h a magnetic clamp Figure 9 pos 2 z z Put the MT strip upon the prepared number platform and fix the MT end on the examined surface using a magnetic clamp avoid warps and shifts relative to the examin...

Page 25: ...spring clip z z Select the CMS concentrator according to the properties of the examined object z z Make a magnetic copy Lower the concentrator of the magnetic scanner unto the fixed MT strip and slig...

Page 26: ...ed object may appear z z Put the magnetic copy into the corresponding cartridge for storing MT see Figures 6a b The one for storing MT with recorded data is marked with red marker If necessary use the...

Page 27: ...ing the cleaning tape wait for at least 3 minutes until the tape drive surfaces get dry before starting device operation Rate of one time ethyl alcohol application 0 5 g Maintenance of the MOV USB dev...

Page 28: ...surface measure the thickness of lacquer and paint coating When examining the stitched image there are image fragments irrelevant to the examined object MT is not completely demagnetized Demagnetize t...

Page 29: ...the button of the pressing roller block It is strictly prohibited to use metal instruments for extracting foreign objects from the tract If the described methods are ineffective it is recommended to a...

Page 30: ...screw for about 1 mm Then put pieces of cardboard on the arrises and tighten plastic straps Figure 12 pos 2 The device is to be stored in climatic conditions given in paragraph 5 Protect the device fr...

Page 31: ...ULA Ltd Address for mailing claims with regard to the quality of the device 1 Volokha Street Office 314 Minsk 220036 Republic of Belarus Phone 375 17 2862825 Fax 375 17 2136897 e mail support regulafo...

Page 32: ...1 Volokha Street Office 314 Minsk 220036 Republic of Belarus Phone 375 17 2862825 Fax 375 17 2136897 e mail mail regula by http www regulaforensics com...

Reviews: