REGULA 7517A User Manual Download Page 28

28

4

4.1

4.2

ROUTINE REPAIRS

General Instructions

The list of possible problems and troubleshooting is given in Table 2.

Table 2

Device

Problem

Cause

Troubleshooting

MOV USB device

When connecting 

the device to the USB 

port of the PC, the 

power indicator does 

not come on.

The USB connector 

has not been 

successfully 

connected.

Check the power cable 

connection: unplug it and then 

plug it in again.

The controller or the 

power cable is out of 

order.

Service center repairs.

Spontaneous MT 

stopping in the OMB 

tape drive tract.

The tape drive is 

clogged with foreign 

objects.

Clean the tape drive as 

described in paragraph 4.2.

MT got stuck in 

the tape drive because 

MT is crumpled or 

damaged.

Straighten or replace the MT 

strip.

Stitching quality 

worsens — images 

appear double.

MT sliding because 

of the tape drive 

clogging.

Clean the tape drive 

mechanism as described in 

paragraph 3.1.

The device does 

not respond to the 

executive commands.

Software error.

Restart the software. 

If necessary, restart the OS.

MCAS

The rubber sleeve 

gasket of the CMS 

rotary concentrator 

is damaged.

Natural wear-and-

tear.

Replace the rubber sleeve 

gasket as described in 

paragraph 4.2.

No precision when 

viewing the stitched 

image

Improper preparation 

of the examined 

surface or a thick 

lacquer-and-paint 

coating.

Improve preparation of 

the examined surface, 

measure the thickness of 

lacquer-and-paint coating.

When examining 

the stitched image, 

there are image 

fragments irrelevant to 

the examined object.

MT is not completely 

demagnetized.

Demagnetize the MT 

strip and carry out a new 

magnetic copy.

When examining 

the stitched image, 

there are longitudinal 

tracks irrelevant to the 

examined object.

Adhesion of 

magnetic particles to 

the scanner surface or 

scratches on the MT 

surface.

Clean the scanner as 

described in paragraph 3.2, 

demagnetize or replace the 

MT and carry out a new 

magnetic copy.

Repair Instructions

Cleaning the OMB tape drive tract if clogged

Cleaning is carried out when the tract is clogged up with foreign objects which interfere with 

MT movement.

The  same  is  applied  to  the  MT  extraction,  when  it  was  jammed  during  the  automatic 

injection or reverse motion. Such jamming is possible when the user does not observe the 

recommended edge form of the magnetic tape strip (see paragraph 2.2.2.2).

To clean the OMB tape drive tract, do the following:

 

{

Turn the front side of the MOV USB device, when in the working position, to 

the operator (see Figure 3).

 

{

Release the MT pressing roller by pressing the button of the pressing roller 

block (Figure  3, pos. 2)

 

{

Still holding the button of the pressing roller block, take the protruding end of 

the MT strip and extract the jammed piece (Figure 11). If two ends of the MT 

stick out, it can be extracted in any direction. In case the length of the tape 

is  less  than  200  mm  or  the  MT  strip  has  rolled  in  the  tape  drive,  none  of 

Summary of Contents for 7517A

Page 1: ...WARE AND SOFTWARE SYSTEM FOR FORENSIC EXAMINATION OF FIREARMS AND AMMUNITION SERIAL NUMBERS REGULA 7517 MAGNETO OPTICAL DEVICE REGULA 7517A USER S GUIDE Certified Quality Management System Regula Ltd...

Page 2: ......

Page 3: ...sories Set MCAS 13 1 2 3 Personal Computer PC 14 1 2 4 Software 15 1 2 5 Spare Parts and Repair Kit SPRK 15 2 INTENDED USE 16 2 1 Device Intended Use 16 2 1 1 Safety measures and operating limitations...

Page 4: ...ssories set MOI magneto optical image MOV magneto optical visualization MOV USB device USB device for magneto optical visualization OMB optic mechanical block MT magnetic tape SPRK spare parts and rep...

Page 5: ...ining and get clearance to perform expert examination in the field of trace evidence z z be an OS Windows user z z study techniques of authenticity verification of firearms and ammunition serial numbe...

Page 6: ...laboratory conditions due to the fact that the device provides high quality results visual evidence and uses non destructive magneto optical methods In case any changes in serial numbers are detected...

Page 7: ...than 5 years Operation warranty period 12 months from the date the device is delivered to the customer Storage warranty period 12 months Dimensions In transportation packaging not more than 350 280 22...

Page 8: ...sories 1 box 1b Peripheral examination tools Regula 7516 body case Regula 7516 Software PC drivers and utilities CD folder CD 1 OS Windows XP CD 2 MS Office CD 3 Special software for managing USB devi...

Page 9: ...where it is automatically detected and introduced into the tape drive tract Then it is scanned step by step by a sensor designed on the basis of a magneto optical crystal Magneto optical visualizatio...

Page 10: ...g with fusible metals dot welding and pressing in fragments See paragraph 2 2 1 2 for the details of the device operational check up using the test object Test object specifications The test object se...

Page 11: ...r s name and or trade mark z z manufacturer s model identification z z device serial number in compliance with the numbering system of the manufacturer z z date of manufacturing year and month z z rat...

Page 12: ...te time of input of a magnetic copy of the number platform 15 sec OS WindowsXPorlater Format of image files BMP PCX GIF JPG PC monitor size 12 Approximate size of an image file of the number platform...

Page 13: ...he copy reflects relief and structural inhomogeneity of the material Permanent magnets of the combination magnetic scanner CMS concentrators serve as sources of magnetic stray fields excitation z z De...

Page 14: ...ced examination of metal structure of objects is the eddy current scanner supplied optionally Its specifications are given in the user s guide to eddy current magnetographing device Regula 7517B Magne...

Page 15: ...ion results Detailed description of NUCA EYER software is given in the corresponding user s guide Spare Parts and Repair Kit SPRK Function SPRK is intended for routine maintenance and repairs of the d...

Page 16: ...z Consider operating limitations for the components of the device in conformity with paragraph 2 2 2 2 1 2 1 1 Failure to observe operating limitations given in paragraphs 2 1 1 and restrictions on th...

Page 17: ...rnal power supply source The device can be powered by the internal autonomous PC battery In this case power supply cables are not connected z z Switch on the PC Start the OS and launch NUCA EYER softw...

Page 18: ...d it will move forward automatically z z Wait for approximately 5 seconds to load the potentially informative sector of the magnetic copy into MOV zone Start the stitching process by corresponding sof...

Page 19: ...mined number Measure geometric dimensions of the images of the digits and analyze their position The software allows measuring linear and angular image dimensions z z Analyze examination results Docum...

Page 20: ...Figure 5a to move the MT strip Figure 5 pos 3 through the slot beyond the edge of the blade Figure 5 pos 2 z z Take the MT by its edge Figure 5 pos 3 and pull it out for a required length relative to...

Page 21: ...netized MT Inserting MT strips into the cartridge and extracting them Figure 6 d c b 1 cartridge for storing MT 2 magnetic clamp 3 MT 1 2 To insert MT strips into the cartridge do the following Roll u...

Page 22: ...irty MT surface greasy film water solvents chemical solutions particles of dust rust metal etc MT surface should be cleaned with alcohol using a special container from the SPRK Figure 7d MT with consi...

Page 23: ...tic MT in the conditions which improve the quality of magnetic copying e g for examining the relief and structure of aluminum objects for examining weak internal stresses in ferromagnetic objects for...

Page 24: ...h a magnetic clamp Figure 9 pos 2 z z Put the MT strip upon the prepared number platform and fix the MT end on the examined surface using a magnetic clamp avoid warps and shifts relative to the examin...

Page 25: ...spring clip z z Select the CMS concentrator according to the properties of the examined object z z Make a magnetic copy Lower the concentrator of the magnetic scanner unto the fixed MT strip and slig...

Page 26: ...ed object may appear z z Put the magnetic copy into the corresponding cartridge for storing MT see Figures 6a b The one for storing MT with recorded data is marked with red marker If necessary use the...

Page 27: ...ing the cleaning tape wait for at least 3 minutes until the tape drive surfaces get dry before starting device operation Rate of one time ethyl alcohol application 0 5 g Maintenance of the MOV USB dev...

Page 28: ...surface measure the thickness of lacquer and paint coating When examining the stitched image there are image fragments irrelevant to the examined object MT is not completely demagnetized Demagnetize t...

Page 29: ...the button of the pressing roller block It is strictly prohibited to use metal instruments for extracting foreign objects from the tract If the described methods are ineffective it is recommended to a...

Page 30: ...screw for about 1 mm Then put pieces of cardboard on the arrises and tighten plastic straps Figure 12 pos 2 The device is to be stored in climatic conditions given in paragraph 5 Protect the device fr...

Page 31: ...ULA Ltd Address for mailing claims with regard to the quality of the device 1 Volokha Street Office 314 Minsk 220036 Republic of Belarus Phone 375 17 2862825 Fax 375 17 2136897 e mail support regulafo...

Page 32: ...1 Volokha Street Office 314 Minsk 220036 Republic of Belarus Phone 375 17 2862825 Fax 375 17 2136897 e mail mail regula by http www regulaforensics com...

Reviews: