6.3 Main Menu
6.3.1 Main Menu Screen
The Touch screen enables quick navigation by
pressing directly on the screen.
e.g. if you would like to change the current
range, press on the text "Range" and the Menu
Range will open. This is a short cut step to the
Range menu. Or access Range menu by going
through the Setup menu.
Start
Starts the measurement and the selected Current is applied.
Stop
Stops the measurement; DO NOT disconnect the measuring cables BEFORE the Green light is on
and the indication Safe appears. This button is active when Current is on (Start is active when
Current is off).
Store
Stores the results while the measuring Current is on.
Demag
Demagnetizes the core of a test object. See chapter 6.3.5
Setup
Goes directly to the Setup Menu. See chapter 6.3.6
Profile
Goes directly in the Profile Menu. See chapter 6.3.13
If you don't use Profiles, you will have the
Result
Button. See chapter 6.3.12
H1-H2
Enable or Disable measuring channel 1 by selecting the small box (next to caption H1-H2).
X1-X2
Enable or Disable measuring channel 2 by selecting the small box (next to caption X1-X2).
5
If you have not selected "Use Profiles" in the Setup menu, you will see "Results" instead of the “Profile” Button
Instruction Manual Winding Resistance Meter
Type WR100 / WR50 / WR14
Page 44 of 139