Serial Bus Analysis
R&S
®
RTC1000
148
User Manual 1335.7352.02 ─ 04
11.7.1
LIN Bus Configuration
Before you configure the bus, set the correct logic level (threshold).
●
●
"To set the threshold for logic states"
default value is 500
mV.
To select the LIN bus
1. Press the [BUS] key in the Vertical section.
2. In the short menu, select the bus: "B1" or "B2".
3. In the "BUS" menu, press the "BUS TYPE" softkey.
4. Use the [Universal] knob to select "LIN":
To configure the CAN bus
1. In the "BUS" menu, press "CONFIGURATION".
2. Configure the bus. The settings are described below.
Figure 11-19: LIN bus configuration
LIN Bus (Option R&S
RTC-K3)