Serial Bus Analysis
R&S
®
RTC1000
142
User Manual 1335.7352.02 ─ 04
At the physical level, CAN is a differential signal, therefore a differential probe is rec-
ommended for decoding. Nevertheless, standard probes are also suitable to capture
the signals. The standard data rates range between 10 kBit/s and 1 MBit/s.
A CAN message primarily consists of a start bit, the frame ID (11 bit or 29 bit), the data
length code DLC, the data, a CRC, acknowledge bit, and an end bit.
11.6.1
CAN Bus Configuration
Before you configure the bus, set the correct logic level (threshold).
●
●
"To set the threshold for logic states"
default value is 500
mV.
To select the CAN bus
1. Press the [BUS] key in the Vertical section.
2. In the short menu, select the bus: "B1" or "B2".
3. In the "BUS" menu, press the "BUS TYPE" softkey.
4. Use the [Universal] knob to select "CAN":
To configure the CAN bus
1. In the "BUS" menu, press "CONFIGURATION".
2. Configure the bus. The settings are described below.
CAN Bus (Option R&S
RTC-K3)