Getting Started
R&S
®
FSVA3000/ R&S
®
FSV3000
83
User Manual 1178.8520.02 ─ 01
(4)
Trace Mode
Sweep mode:
Clrw
CLEAR/WRITE
Max
MAX HOLD
Min
MIN HOLD
Avg
AVERAGE (Lin/Log/Pwr)
View
VIEW
(5)
Smoothing factor
Smth
Smoothing factor, if enabled.
(See
Norm/NCor
Correction data is not used.
5.4.1.3
Marker Information
Marker information is provided either in the diagram grid or in a separate marker table,
depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x-axis and y-axis positions of the last two markers or delta
markers that were set are displayed, if available, as well as their index. The value in
the square brackets after the index indicates the trace to which the marker is assigned.
(Example: M2[1] defines marker 2 on trace 1.) For more than two markers, a separate
marker table is displayed beneath the diagram by default.
Marker information in marker table
In addition to the marker information displayed within the diagram grid, a separate
marker table may be displayed beneath the diagram. This table provides the following
information for all active markers:
Type
Marker type: N (normal), D (delta), T (temporary, internal), PWR (power sensor)
Ref
Reference (for delta markers)
Trc
Trace to which the marker is assigned
X-value
X-value of the marker
Y-value
Y-value of the marker
Func
Activated marker or measurement function
Func .Result
Result of the active marker or measurement function
The functions are indicated with the following abbreviations:
FXD
Fixed reference marker
PHNoise
Phase noise measurement
Operating the Instrument