Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
921
User Manual 1178.8520.02 ─ 01
See
"Adjusting settings automatically during triggered measurements"
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Manual operation:
See
" Adjusting the Center Frequency Automatically ( Auto Fre-
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
FSV/A or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
Manual operation:
See
" Setting the Reference Level Automatically ( Auto Level )"
14.7.6
Configuring the Data Input and Output
The following commands are required to configure data input and output.
................................................................................... 930
................................................................................ 941
........................................................................................ 951
14.7.6.1
RF Input
.......................................................................................... 922
INPut<ip>:ATTenuation:PROTection:RESet
.......................................................................922
..........................................................................................923
Setting Basic Measurement Parameters