Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
930
User Manual 1178.8520.02 ─ 01
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<Type>
String containing one of the following values:
–
"None" (no probe detected)
–
"active differential"
–
"active single-ended"
–
"active modular"
Example:
//Query probe type
PROB3:SET:TYPE?
Usage:
Query only
Manual operation:
See
14.7.6.3
External Generator Control
External generator control commands are available if the R&S
FSV/A External Genera-
tor Control option (R&S
FSV3-B10) is installed.
For each measurement channel one external generator can be configured. To switch
between different configurations define multiple measurement channels.
For more information on external generator control see
..................................................................................930
..........................................................................................934
.................................................................................................936
Programming Example for External Generator Control
........................................ 939
Measurement Configuration
The following commands are required to activate external generator control and to con-
figure a calibration measurement with an external tracking generator.
SOURce<si>:EXTernal<gen>:FREQuency
........................................................................930
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
............................................ 931
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:DENominator
......................................931
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:NUMerator
......................................... 932
SOURce<si>:EXTernal<gen>:FREQuency:OFFSet
........................................................... 932
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
................................................................... 933
SOURce<si>:EXTernal<gen>[:STATe]
.............................................................................. 933
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
.............................................................934
SOURce<si>:EXTernal<gen>:FREQuency
<Frequency>
This command defines a fixed source frequency for the external generator.
Suffix:
<si>
.
irrelevant
Setting Basic Measurement Parameters